Circuit Parameter 7 | Measure internal resistance. Why do we have four wires?

Circuit Parameter 7 | Measure internal resistance. Why do we have four wires?

In the previous chapter, we distinguished AC internal resistance and DC internal resistance. However, even if the test method is the same, the reading may change if you change a wire and re-pin the coil.
One reason for this is that: The device may have also factored in the resistance of the wires and the contact points into its results.
In this chapter, we’ll learn about the “ four‑line method ” for reducing such effects.

01 | The internal resistance of the electrical core is small, the external resistance cannot be ignored

Suppose that in a test, the equivalent internal resistance of the electrical core under the prescribed conditions is 0.30 μm, and the combined contribution of the two side wires and contact position is 0.20 μm.
If the measurement included them all, the result might be close to:

0.30 + 0.20 = 0.50 MEU.

The electrical core did not change, but the reading was significantly larger.
This is just a simplified example of the source of the error. Real-world testing is also influenced by factors such as temperature, load state and instrumentation.

02 | Two-line method: The same pair of wires, both carrying current and measuring voltage

To understand internal impedance measurement, one can grasp a relationship:

Apply current change → Measure voltage response → Calculate internal resistance.

In the two-wire connection, current transmission and voltage measurement share a pair of conductors.
When the current passes through the contact position of the wire and the clip, it also produces a pressure drop. If the voltage sampling contains these locations, the calculation results are mixed with external resistors.
So the problem is not just whether the “wire is thick enough”; it also involves where exactly the voltage is measured .

03 | Four-line method: separating the current traveling from the measured voltage

The four‑wire method, also known as Kelvin wiring (Kelvin connection) ,Divide the connections into two groups:
Connect
Effect
Two current lines
Apply a test current to the core or assume a charge and discharge current
Two voltage sampling lines
Direct detection of specified voltage between sampling points
The positive and negative poles of the electrical core connect a current line and a sampling line each.
It can be understood colloquially as:

The current line is responsible for "doing the test," and the sampling line is concerned with "seeing what happens at both ends of the coil."

The sampling line should be connected to the specified location on the battery cell’s terminal – not arbitrarily attached to a device terminal that is far away from the battery cell. Hichi : The principle of four‑terminal measurement

04 | The sampling line also has resistance. Why does it have a small effect?

The key point is this: The input impedance at the voltage‑measurement terminal is very high, so the current flowing through the sampling line is extremely small.
According to the "pressure drop = current × resistance," even if the sampling line has an resistance, the pressure drop itself is usually small.
At the same time, the pressure drop on the current line is located outside the selected voltage sampling point and is no longer directly counted in the measured voltage response.
The four-wire method does not make the wire resistance disappear, but instead reduces its interference with the result by separating the path.

05 | When you connect four lines, is it always accurate?

It's not enough. The location of the sampling point and the state of contact are equally important.
If the sampling line is connected to the busbar on the outside, then the connection resistance between the sampling points may still be included。The surface condition of the terminals, the clamping position, and the current distribution may also affect the readings.
Therefore, when comparing the same batch of cells, it should be as uniform as possible:
Sampling location and method of contact of the jig;
Temperature, load state and static condition;
The alternating current frequency, or direct‑current pulses and the sampling time.
In low resistance measurement, probe arrangement is itself part of the measurement method. Day-set: probe arrangement for low resistance measurement
Also distinguish between: * * Four-line method describes the connection method, and AC and DC methods describe the incentive and calculation method. * * The use of a four-wire connection does not make the AC internal resistance and DC internal resistance the same number.
Remember a sentence in this chapter

To measure internal resistance, we must not only ask "how to measure it," but also ask "what is contained between the voltage sampling points."